We use cookies to understand how you use our site and to improve your experience. This includes personalizing content and advertising. To learn more, click here. By continuing to use our site, you accept our use of cookies. Cookie Policy.

Features Partner Sites Information LinkXpress hp
Sign In
Advertise with Us
GLOBETECH PUBLISHING LLC

Download Mobile App




US FDA Issues Class 2 Device Recall of CT Scanners

By MedImaging International staff writers
Posted on 14 Apr 2015
The US Food and Drug Administration (FDA) issued a Class 2 Recall notice on April 1, 2015, for several Philips Healthcare (Best, The Netherlands) Computed Tomography (CT) scanners worldwide.

According to the FDA, “Philips discovered that a software defect exists in the marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.”

The CT systems affected are the Philips Healthcare Brilliance CT 64-channel scanner with Essence technology, the Brilliance iCT, the Brilliance iCT SP, the Ingenuity Core, the Ingenuity Core128, and the Ingenuity CT scanner.

Philips Healthcare already sent a customer information letter, dated February 5, 2015, to all affected customers. The letter identified the products affected, the problem found, and the action that the customer needed to take. The problem was determined to be in the software design, and Philips Healthcare informed their customers to refrain from using Multi Planar Reconstruction (MPR) derived images for identifying the absolute position of the intervention for feet-first patient orientation.

To correct the issue, Philips Healthcare has instructed field service engineers to contact affected customers and schedule on-site software upgrade installations.

Related Links:

Philips Healthcare
US Food and Drug Administration (FDA)


Pocket Fetal Doppler
CONTEC10C/CL
Ultrasound Table
Women’s Ultrasound EA Table
Mammography System (Analog)
MAM VENUS
New
Floor‑Mounted Digital X‑Ray System
MasteRad MX30+

Channels

Nuclear Medicine

view channel
Image: The diagnostic tool could improve diagnosis and treatment decisions for patients with chronic lung infections (Photo courtesy of SNMMI)

Novel Bacteria-Specific PET Imaging Approach Detects Hard-To-Diagnose Lung Infections

Mycobacteroides abscessus is a rapidly growing mycobacteria that primarily affects immunocompromised patients and those with underlying lung diseases, such as cystic fibrosis or chronic obstructive pulmonary... Read more

Imaging IT

view channel
Image: The new Medical Imaging Suite makes healthcare imaging data more accessible, interoperable and useful (Photo courtesy of Google Cloud)

New Google Cloud Medical Imaging Suite Makes Imaging Healthcare Data More Accessible

Medical imaging is a critical tool used to diagnose patients, and there are billions of medical images scanned globally each year. Imaging data accounts for about 90% of all healthcare data1 and, until... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.